Lot nr. 97 Sistem radiografic digital (cu bucky vertical) cerințe medii 210211
Information about customer
Fiscal code/IDNO
Address
MD-2005, MOLDOVA, mun.Chişinău, mun.Chişinău, mun. Chișinău MD-2005, bd. Grigore Vieru, 22/2
Web site
---
The contact person
Full name
Aurelia Botica
Contact phone
022222364
Purchase data
Date created
22 Jul 2024, 14:53
Date modified
5 Sep 2024, 18:02
Estimated value (without VAT)
12 000 000 MDL
Achizitii.md ID
21261412
Type of procedure
Open tender
Award criteria
The lowest price
Delivery address
MD-2005, MOLDOVA, mun.Chişinău, mun.Chişinău, mun. Chișinău MD-2005, bd. Grigore Vieru, 22/2
Contract period
1 Jan 2025 02:00 - 30 Dec 2027 02:00
List of positions
1)
Title
Lot nr. 97 Sistem radiografic digital (cu bucky vertical) cerințe medii 210211 CPV: 33100000-1 - Medical equipments
Quantity: 8.0
Unit of measurement: Bucata
Documents of the procurement procedure
ds_bunuri_servicii_omf_115_15_09_2021.docx
Bidding Documents
-
22.07.24 14:58
duae_ro_0 (1).doc
Bidding Documents
-
22.07.24 14:58
contract nou.docx
Bidding Documents
-
22.07.24 14:58
acord - cadru 1.doc
Bidding Documents
-
22.07.24 14:58
formularele (3).xlsx
Bidding Documents
-
5.09.24 18:02
Lot nr. 97 Sistem radiografic digital (cu bucky vertical) cerințe medii 210211
Date:
24 Jul 2024, 10:31
Question's name:
Clarificari
Question:
- Mișcarea mesei Laterală, cm ± 13: intrucat unii producatori renumit au miscarea laterala mai mica, solicitam modificarea in: Mișcarea mesei Laterală, cm ≥ ± 12,5 - Panou de control pe tubul radiologic obligatoriu: intrucat se solicita un system mediu, solicitam excluderea acestei cerinta sau modificarea in optional - Suspensia tubului Instalare/fixare podea tip motorizat obligatoriu: acest parametru este characteristic sistemelor avansate. Solicitam excluderea. - Aria de scanare a pacientrului ≥ 180 cm: acest parametru pare confuz. Aria de scanare este echivalenta cu distanta de deplasare al sistemului Bucky. Acest parametru ar trebui modificat in: Distanta de deplasare a tubului deasupra mesei ≥ 180 cm - Sincronizare automata cu tubul radiologic obligatoriu:
Answer (19 Aug 2024, 14:18):
: Se acceptă modificarea cerinței în „Mișcarea mesei / Laterală, cm minim ± 12.5”. Se acceptă modificarea cerinței în „Suspensia tubului / Panou de control pe tubul radiologic - opțional”. Se acceptă reformularea cerinței în „Distanta de deplasare a tubului deasupra mesei”.
Unfortunately, questions can be asked only during the "Active" period.