1
Enquiry period
with 05.05.2020 15:02
to 11.05.2020 10:00
2
Bidding period
with 11.05.2020 10:00
to 14.05.2020 10:00
3
Auction
with 15.05.2020 15:00
to 15.05.2020 15:23
4
Evaluation

5
Contract

Status Evaluation
Estimated value without VAT 731 333 MDL
Period of clarifications: 5 May 2020, 15:02 - 11 May 2020, 10:00
Submission of proposals: 11 May 2020, 10:00 - 14 May 2020, 10:00

Supplier technical support:

(+373) 79999801

Aparate de ventilare artificială a plămînilor

Information about customer
Fiscal code/IDNO
Address
3401, MOLDOVA, Hînceşti, mun.Hînceşti (r-l Hînceşti), Str. M. Nincu, 126
Web site
---
The contact person
Full name
Lupascu Simion
Contact phone
068101202
Purchase data
Date created
5 May 2020, 15:02
Date modified
8 May 2020, 12:27
Achizitii.md ID
21022499
CPV
33100000-1 - Echipamente medicale
Type of procedure
Request price offers (goods)
Award criteria
The lowest price
Funding sources
List of lots
Documents of the procurement procedure
aparate ventilare.pdf aparate ventilare.pdf
tenderNotice
4 file
5.05.20 15:02
Aparate de ventilare artificială a plămînilor
Date:
8 May 2020, 12:27
Question's name:
Regimuri de ventilare
Question:
Cum va avea loc evaluarea regimurilor de ventilare? Strict conform marcarilor indicate DouPAP, SPONT sau se accepta si analogicul lor (DouPAP - VCV (Volume Control Ventilation), SPONT - BPAP(Bilevel Positive Airway Pressure))? Regimul DouPAP, SPONT este specifica unui singur producator Hamilton. In caz ca nu se accepta rugam modificarea cerintelor in sintagma DouPAP sau analogic si SPONT sau analogic.
Answer (11 May 2020, 09:33):
Evaluarea regimurilor de ventilare se va efectua in comparatie. In cazul analogului regimului solicitat el va fi luat in considerare, dar cu o descriere completa si corespunderea unei calitati conform regimurilor solicitate.
Aparate de ventilare artificială a plămînilor
Date:
8 May 2020, 12:41
Question's name:
Regim de ventilare ASV
Question:
Regimul de ventilare ASV (Adaptive Support Ventilation) este specific unui singur producator (Hamilton). Rugam respectuos eliminarea cerintei sau modificarea in VS(Volume Support) sau PS(Pressure Support) sau PPS(Proportional Pressure Support), pentru posibilitatea de participare a mai multor agenti economici.
Answer (11 May 2020, 09:36):
Se permite si indicarea altor sintagme, dar ca rezultat final sa fie prezent regimul adaptiv de ventilare obligatoriu.
Aparate de ventilare artificială a plămînilor
Date:
8 May 2020, 13:14
Question's name:
Fluxul maxim
Question:
Cerinta „Fluxul maxim de inspirație 240 l/min (150 l/ min cu 100 % O2)” este un parametru motamo din datele tehnice al produsului de la Hamilton - „Maximum inspiratory flow 240 l/min (150 l/min with 100% O2)”. Rugam respectuor excluderea acestei cerinte pentru a nu limita ceilalti participanti.
Answer (11 May 2020, 09:41):
Fluxul de inspirație este unul din parametrii principali. Nu este un parametru motamo, este un parametru pentru dispozitiv din clasa de ventilare pulmonara de calitate. Acest parametru poate fi regasit si la Drager de-o pilda, precum si la alte modele.
Aparate de ventilare artificială a plămînilor
Date:
8 May 2020, 13:18
Question's name:
TI – 0.1-12 sec
Question:
Rugam modificarea parametrului TI – 0.1-12 sec in minim 0.2-8 sec. acesta fiind pe deplin sufiucient pentru utilizarea normala a dispozitivului de ventilare si va permite marei majoritati a produselor de a participa la concurs.
Answer (11 May 2020, 09:46):
TI – 0.1-12 sec nu va fi modificat, Se admite ca parametrul sa fie cu valori mai ridicate de valoarea maxima de 12 sec. si nici de cum cu valori mai mici. Se doreste un dispozitiv, care va fi necesar nu pentru utilizarea normala, dar pentru ventilarea artificiala de calitate